RegWardenFDA facility compliance records
Recall

Class II — TERRAGENE S.A.

Recalling firmTerragene SA
Registered facilityTERRAGENE SA (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonError in the IFU lists an incorrect incubation time on one occasion of 20 minutes rather than the correct incubation time of 60 minutes.
ProductASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.
Quantity914 devices in the U.S.
DistributionUS Nationwide distribution in the states of CA and KY.
Initiated2025-01-15
Classified2025-02-27
Recall numberZ-1248-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — TERRAGENE S.A. (Z-1248-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96176/