RegWardenFDA facility compliance records
Recall

Class II — HUMAN MED AG

Recalling firmHuman Med AG
Registered facilityHUMAN MED AG (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
Producthuman med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
Quantity775 devices
DistributionUS Nationwide distribution in the state of FL.
Initiated2024-10-18
Classified2025-02-28
Recall numberZ-1251-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — HUMAN MED AG (Z-1251-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96137/