Class II — Aerin Medical, Inc.
| Recalling firm | Aerin Medical Inc. |
|---|---|
| Registered facility | AERIN MEDICAL INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console. |
| Product | Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile. |
| Quantity | 479 units |
| Distribution | US Nationwide Distribution: AL, AZ, CA, CO, GA, ID, IL, IN, KY LA, MI, MN, NY, NE, NY, OH, OR PA, SC, TN, TX, UT, VA, WA, WI, and WV. |
| Initiated | 2024-11-12 |
| Classified | 2025-01-06 |
| Recall number | Z-0808-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Aerin Medical, Inc. (Z-0808-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95832/