Class III — Qualigen Inc

Recalling firmQUALIGEN, INC.
Registered facilityQUALIGEN, INC. (the firm's only registered facility)
ClassificationClass III
StatusOngoing
ReasonThree lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
ProductThe FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.
Quantity426
DistributionDistribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.
Initiated2024-12-03
Classified2025-01-03
Recall numberZ-0805-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class III recall — Qualigen Inc (Z-0805-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95807/

Not open yet

Alerts for QUALIGEN, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.