RegWardenFDA facility compliance records
Recall

Class II — LivaNova Deutschland GmbH

Recalling firmLivaNova Deutschland GmbH
Registered facilityLivaNova Deutschland GmbH (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonHeart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
ProductEssenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Quantity38
DistributionUS Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.
Initiated2024-10-18
Classified2024-11-27
Recall numberZ-0571-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — LivaNova Deutschland GmbH (Z-0571-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95665/

Not open yet

Alerts for LivaNova Deutschland GmbH

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.