Class II — Roche Diagnostics Operations, Inc.
| Recalling firm | Roche Diagnostics Operations Inc. |
|---|---|
| Registered facility | View this company's 14 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias) |
| Product | cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine |
| Quantity | 458 units |
| Distribution | US: NJ, TX, NY, FL, IN, TX, OK, SC, IA, |
| Initiated | 2024-10-17 |
| Classified | 2024-11-19 |
| Recall number | Z-0489-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Roche Diagnostics Operations, Inc. (Z-0489-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95609/