RegWardenFDA facility compliance records
Recall

Class II — Micro-X Ltd.

Recalling firmMICRO-X LTD
Registered facilityMICRO-X LTD (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonShots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
ProductMICRO-X Rover Mobile X-ray System, # MXU-RV35
Quantity5
DistributionWorldwide and US Nationwide Distribution: NJ, PR, CA
Initiated2023-05-31
Classified2024-12-12
Recall numberZ-0664-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Micro-X Ltd. (Z-0664-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95603/