Class II — Defibtech, LLC

Recalling firmDEFIBTECH, LLC
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonWithin the Irish Market, a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life. There is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient.
ProductDefibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XX Series)- Indicated for use on victims of sudden cardiac arrest (SCA)
Quantity22,294 total
DistributionIreland
Initiated2024-10-14
Classified2024-11-21
Recall numberZ-0502-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Defibtech, LLC (Z-0502-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95594/

Not open yet

Alerts for DEFIBTECH, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.