Class II — Siemens Healthcare Diagnostics, Inc.
| Recalling firm | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU). |
| Product | IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282) |
| Quantity | 7,332 units |
| Distribution | Worldwide - US Nationwide distribution. |
| Initiated | 2024-09-30 |
| Classified | 2024-11-14 |
| Recall number | Z-0464-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-0464-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95513/
Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.