Recall

Class II — Accriva Diagnostics, Inc.

Recalling firmAccriva Diagnostics, Inc.
Registered facilityAccriva Diagnostics, Inc. (the firm's only registered facility)
ClassificationClass II
StatusCompleted
ReasonValues used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.
ProductAVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.
Quantity1
DistributionUS: AZ
Initiated2024-09-27
Classified2024-10-31
Recall numberZ-0279-2025
InitiationVoluntary: Firm initiated