Class II — Accriva Diagnostics, Inc.
| Recalling firm | Accriva Diagnostics, Inc. |
|---|---|
| Registered facility | Accriva Diagnostics, Inc. (the firm's only registered facility) |
| Classification | Class II |
| Status | Completed |
| Reason | Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention. |
| Product | AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter. |
| Quantity | 1 |
| Distribution | US: AZ |
| Initiated | 2024-09-27 |
| Classified | 2024-10-31 |
| Recall number | Z-0279-2025 |
| Initiation | Voluntary: Firm initiated |