Class II — FUJIFILM Healthcare Americas Corporation

Recalling firmFUJIFILM Healthcare Americas Corporation
Registered facilityFUJIFILM Healthcare Americas Corporation (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonWhen the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
ProductThe device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
Quantity130
DistributionUS Nationwide and Worldwide
Initiated2023-11-23
Classified2024-09-25
Recall numberZ-3158-2024
InitiationFDA Mandated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — FUJIFILM Healthcare Americas Corporation (Z-3158-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95320/

Not open yet

Alerts for FUJIFILM Healthcare Americas Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.