RegWardenFDA facility compliance records
Recall

Class II — Healgen Scientific Llc

Recalling firmHEALGEN SCIENTIFIC LLC
Registered facilityHEALGEN SCIENTIFIC LLC (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.
ProductHealgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.
Quantity1,636,994 Tests
DistributionU.S. Nationwide distribution in the states of CA, GA and TX.
Initiated2024-08-22
Classified2024-09-27
Recall numberZ-3303-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Healgen Scientific Llc (Z-3303-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/95248/

Not open yet

Alerts for HEALGEN SCIENTIFIC LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.