Class I — Breas Medical, Inc.

Recalling firmBREAS MEDICAL, INC.
Registered facilityBREAS MEDICAL, INC. (the firm's only registered facility)
ClassificationClass I
StatusOngoing
ReasonThere is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in specific conditions.
ProductVivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory support for patients who require invasive or non-invasive mechanical ventilation, Model/Catalog Number: 230000 (230016, 230216)
Quantity8298
DistributionUS Nationwide.
Initiated2024-07-25
Classified2024-09-03
Recall numberZ-2690-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)