Recall

Class II — Beckman Coulter, Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThe reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
ProductBeckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Quantity183 devices
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, and TX. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
Initiated2024-05-22
Classified2024-08-09
Recall numberZ-2544-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter, Inc. (Z-2544-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94841/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.