Class II — B. Braun Medical, Inc.
| Recalling firm | B. BRAUN MEDICAL INC. |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood. |
| Product | Infusomat SPACE PUMP IV SET W/2 CARESITE, CKVLV -Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490103 |
| Quantity | 643608 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| Initiated | 2024-06-17 |
| Classified | 2024-07-17 |
| Recall number | Z-2400-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — B. Braun Medical, Inc. (Z-2400-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94823/