Class II — Siemens Healthcare Diagnostics Inc
| Recalling firm | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions. |
| Product | Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ; |
| Quantity | 9 units US |
| Distribution | Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and the countries of Chile, Costa Rica, France, Germany, Japan, Lebanon, Mexico, Netherlands, New Zealand, Poland, Spain, Turkey, United Kingdom. |
| Initiated | 2024-05-17 |
| Classified | 2024-06-26 |
| Recall number | Z-2214-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Healthcare Diagnostics Inc (Z-2214-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94704/
Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.