RegWardenFDA facility compliance records
Recall

Class II — Merit Medical Systems, Inc.

Recalling firmMerit Medical Systems, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonDue to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body.
ProductPrelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS -PG6F45MP -PG6F45R -PG6F45S -PG6F45S018 -PG6F65MP -PG6F65S -PG6F65SX -PG6F90MP -PG6F90MPX -PG6F90SX -PG6F90SX018 -PG7F45MP -PG7F45MPX -PG7F45RX -PG7F45S -PG7F45SX -PG7F65SX -PG7F90SX -PG8F45MPX -PG8F45S -PG8F45SX -PG8F65SX -PG8F90SX -X9PG700001 The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.
Quantity52,410 (updated 06/14/2024)
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Brazil, Canada, Cayman Island, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Hong Kong, Republi…
Initiated2024-05-22
Classified2024-06-27
Recall numberZ-2218-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Merit Medical Systems, Inc. (Z-2218-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94684/

Not open yet

Alerts for Merit Medical Systems, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.