Class II — Merit Medical Systems, Inc.
| Recalling firm | Merit Medical Systems, Inc. |
|---|---|
| Registered facility | View this company's 15 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body. |
| Product | Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS -PG6F45MP -PG6F45R -PG6F45S -PG6F45S018 -PG6F65MP -PG6F65S -PG6F65SX -PG6F90MP -PG6F90MPX -PG6F90SX -PG6F90SX018 -PG7F45MP -PG7F45MPX -PG7F45RX -PG7F45S -PG7F45SX -PG7F65SX -PG7F90SX -PG8F45MPX -PG8F45S -PG8F45SX -PG8F65SX -PG8F90SX -X9PG700001 The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature. |
| Quantity | 52,410 (updated 06/14/2024) |
| Distribution | Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Brazil, Canada, Cayman Island, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Hong Kong, Republi… |
| Initiated | 2024-05-22 |
| Classified | 2024-06-27 |
| Recall number | Z-2218-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Merit Medical Systems, Inc. (Z-2218-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94684/
Alerts for Merit Medical Systems, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.