RegWardenFDA facility compliance records
Recall

Class II — Johnson & Johnson Surgical Vision, Inc.

Recalling firmJohnson & Johnson Surgical Vision, Inc.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonDue to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
ProductJohnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
Quantity3,428 units
DistributionU.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.
Initiated2024-03-14
Classified2024-05-02
Recall numberZ-1755-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Johnson & Johnson Surgical Vision, Inc. (Z-1755-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94231/

Not open yet

Alerts for Johnson & Johnson Surgical Vision, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.