RegWardenFDA facility compliance records
Recall

Class II — Navinetics Inc

Recalling firmNaviNetics Inc.
Registered facilityNAVINETICS INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDefective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.
ProductNaviNetics Skull Anchor Key and Drill Kit, REF NN1215
Quantity20 kits
DistributionUS Nationwide distribution in the states of FL, MN, CO.
Initiated2024-02-26
Classified2024-04-04
Recall numberZ-1478-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Navinetics Inc (Z-1478-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94163/

Not open yet

Alerts for NaviNetics Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.