Class II — PHILIPS MEDICAL SYSTEMS
| Recalling firm | Philips Medical Systems International BV |
|---|---|
| Registered facility | View this company's 54 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size. |
| Product | CombiDiagnost R90 1.0 (709030) |
| Quantity | 86 units |
| Distribution | US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA. |
| Initiated | 2022-02-11 |
| Classified | 2024-02-14 |
| Recall number | Z-1087-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)