RegWardenFDA facility compliance records
Recall

Class II — MEDLINE INDUSTRIES, LP - Northfield

Recalling firmMedline Industries, LP
Registered facilityView this company's 31 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonPackaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).
ProductMedline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe
Quantity4831 units
DistributionUS, Nationwide
Initiated2024-01-04
Classified2024-02-16
Recall numberZ-1157-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — MEDLINE INDUSTRIES, LP - Northfield (Z-1157-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93795/