Class II — XTANT Medical Holdings, Inc

Recalling firmXtant Medical Holdings, Inc.
Registered facilityXtant Medical Holdings, Inc. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonSpinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
ProductCortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.
Quantity6
DistributionUS distribution in MI.
Initiated2023-11-28
Classified2024-01-29
Recall numberZ-0846-2024
InitiationVoluntary: Firm initiated