RegWardenFDA facility compliance records
Recall

Class II — SEDECAL SA

Recalling firmSEDECAL S.A.
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonWhile cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
ProductMobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007
Quantity851 devices
DistributionUS Nationwide distribution. There was government distribution but no military distribution.
Initiated2021-10-14
Classified2024-02-14
Recall numberZ-1083-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — SEDECAL SA (Z-1083-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93651/