RegWardenFDA facility compliance records
Recall

Class II — Randox Laboratories Ltd.

Recalling firmRANDOX LABORATORIES, LTD.
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonCK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
ProductCalibration Serum Level 3 CAL2351
Quantity269 kits
DistributionUS Nationwide and Puerto Rico.
Initiated2023-11-20
Classified2024-01-18
Recall numberZ-0752-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Randox Laboratories Ltd. (Z-0752-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93612/

Not open yet

Alerts for RANDOX LABORATORIES, LTD.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.