RegWardenFDA facility compliance records
Recall

Class II — Phadia Ab

Recalling firmPHADIA AB
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonComplaints that specific samples produced false positive EliA GBM results. A positive signal was present when these samples were tested for coating-solution reactivity using EliA wells without antigen.
ProductEliA GBM Wells, Material Number 14551401, for Phadia 250, 2500, and 5000. In vitro diagnostic
Quantity124,928 kits
DistributionUS Nationwide.
Initiated2023-05-24
Classified2024-01-24
Recall numberZ-0783-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Phadia Ab (Z-0783-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93403/