Recall
Class II — Abbott
| Recalling firm | ABBOTT LABORATORIES |
|---|---|
| Registered facility | View this company's 64 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure. |
| Product | Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator |
| Quantity | 34 devices |
| Distribution | US, Taiwan, and Colombia |
| Initiated | 2023-09-28 |
| Classified | 2023-11-02 |
| Recall number | Z-0209-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Abbott (Z-0209-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93152/