RegWardenFDA facility compliance records
Recall

Class II — TECHNO-PATH MANUFACTURING LTD.

Recalling firmTechno-path Manufacturing Ltd.
Registered facilityTECHNO-PATH MANUFACTURING LTD. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonTechnopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
ProductMultichem P, Part Code 08P90-10
Quantity2964 units
DistributionUS Nationwide distribution in the state of Illinois.
Initiated2023-08-25
Classified2023-10-12
Recall numberZ-0106-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — TECHNO-PATH MANUFACTURING LTD. (Z-0106-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93017/

Not open yet

Alerts for Techno-path Manufacturing Ltd.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.