Class II — MEDLINE INDUSTRIES, LP - Northfield
| Recalling firm | Medline Industries, LP |
|---|---|
| Registered facility | View this company's 31 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | 4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user. |
| Product | MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049 |
| Quantity | 324,451 devices |
| Distribution | Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Hong Kong, Taiwan, Singapore |
| Initiated | 2023-08-03 |
| Classified | 2023-09-18 |
| Recall number | Z-2595-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — MEDLINE INDUSTRIES, LP - Northfield (Z-2595-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92881/
Alerts for Medline Industries, LP
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.