RegWardenFDA facility compliance records
Recall

Class II — B. Braun Medical, Inc.

Recalling firmB. BRAUN MEDICAL INC.
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonKits were assembled with the incorrect Filter Straw.
ProductPerifix® / Epidural anesthesia kit (10 count carton)
Quantity16,070 units
DistributionUS Nationwide distribution.
Initiated2023-06-05
Classified2023-06-23
Recall numberZ-2002-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — B. Braun Medical, Inc. (Z-2002-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92523/

Not open yet

Alerts for B. BRAUN MEDICAL INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.