RegWardenFDA facility compliance records
Recall

Class II — Abiomed, Inc.

Recalling firmABIOMED, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThere was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach.
ProductAbiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Quantity37 units
DistributionUS Nationwide distribution in the states of MN, NJ, TN.
Initiated2023-05-26
Classified2023-07-11
Recall numberZ-2128-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)