Class II — Abiomed, Inc.
| Recalling firm | ABIOMED, INC. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach. |
| Product | Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457 |
| Quantity | 37 units |
| Distribution | US Nationwide distribution in the states of MN, NJ, TN. |
| Initiated | 2023-05-26 |
| Classified | 2023-07-11 |
| Recall number | Z-2128-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Abiomed, Inc. (Z-2128-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92521/