Class II — Maquet Cardiovascular, LLC
| Recalling firm | Maquet Cardiovascular LLC |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion. |
| Product | ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest stabilization system for cardiac and chest surgery. |
| Quantity | 785 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates. |
| Initiated | 2023-06-28 |
| Classified | 2023-07-31 |
| Recall number | Z-2271-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Maquet Cardiovascular, LLC (Z-2271-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92446/
Alerts for Maquet Cardiovascular LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.