RegWardenFDA facility compliance records
Recall

Class II — SenTec AG

Recalling firmSentec, Inc.
Registered facilitySENTEC, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.
ProductV-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).
Quantity11 in the U.S.
DistributionUS Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.
Initiated2023-05-02
Classified2023-06-20
Recall numberZ-1980-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — SenTec AG (Z-1980-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92355/