Class II — OmniCell, Inc.
| Recalling firm | Omnicell, Inc. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling |
| Product | Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373 |
| Quantity | 6 units |
| Distribution | AL, NY |
| Initiated | 2023-03-27 |
| Classified | 2023-05-22 |
| Recall number | Z-1617-2023 |
| Initiation | Voluntary: Firm initiated |