RegWardenFDA facility compliance records
Recall

Class II — Richard Wolf GmbH

Recalling firmRICHARD WOLF GMBH
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonPotentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
ProductFLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
Quantity342 units
DistributionWorldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czech Republic (CZ), Germany (DE), Denmark (DK), Ecuador (EC), Spain (ES), Finland (FI), France (FR), Greece (GR), Hong Kong (HK), Croatia (HR), Ireland (IE), India (IN), Italy (IT), Jordan (JO), Japan (JP), South Korea (KR), K…
Initiated2023-03-24
Classified2023-04-24
Recall numberZ-1423-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Richard Wolf GmbH (Z-1423-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91953/

Not open yet

Alerts for RICHARD WOLF GMBH

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.