RegWardenFDA facility compliance records
Recall

Class II — Spacelabs Healthcare, Inc.

Recalling firmSPACELABS HEALTHCARE, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonCommand modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
ProductUltraview SL (UVSL) Command Module, Model 91496
Quantity1,088
DistributionUS Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.
Initiated2023-04-12
Classified2023-05-03
Recall numberZ-1503-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Spacelabs Healthcare, Inc. (Z-1503-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91851/

Not open yet

Alerts for SPACELABS HEALTHCARE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.