Class II — Stryker GmbH
| Recalling firm | STRYKER CORP. |
|---|---|
| Registered facility | View this company's 61 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod. |
| Product | stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved |
| Quantity | 488 units |
| Distribution | US Nationwide distribution. |
| Initiated | 2023-02-07 |
| Classified | 2023-03-01 |
| Recall number | Z-1191-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Stryker GmbH (Z-1191-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91666/
Alerts for STRYKER CORP.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.