RegWardenFDA facility compliance records
Recall

Class II — Coltene Whaledent Inc

Recalling firmColtene/Whaledent AG
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonScrew connection between the turbine head and the push button may loosen and fall off during the treatment, and increase the risk of swallowing or inhaling part of the turbines head or the push button
ProductSciCan STATIS SL S10001 Dental Handpieces-Intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations and for polishing teeth. Model Number: S10001
Quantity3896 units
DistributionNationwide Foreign: Canada Germany Indonesia Ireland Lebanon Saudia Arabia Turkey
Initiated2023-01-20
Classified2023-03-02
Recall numberZ-1208-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Coltene Whaledent Inc (Z-1208-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91601/

Not open yet

Alerts for Coltene/Whaledent AG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.