RegWardenFDA facility compliance records
Recall

Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonPotential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe increased imprecision with low level patient samples
ProductIMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/ IMMULITE 2000 XPi) for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # L2KTY2(D) Siemens Material Number (SMN): 10381647
Quantity717 kits
DistributionNationwide Foreign: Argentina® Australia® Austria® Bahrain® Bangladesh® Belgium® Bolivia® BosniaHerzeg. Brazil® Bulgaria® Canada® Chile® Colombia® CostaRica® Croatia® Cyprus® CzechRepublic® DominicanRep. Ecuador® Egypt® Estonia® Finland® France® Georgia®…
Initiated2022-12-07
Classified2023-01-11
Recall numberZ-0949-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-0949-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91310/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.