Class II — Johnson & Johnson Surgical Vision, Inc.

Recalling firmJohnson & Johnson Surgical Vision, Inc.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonWeld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
ProductFUSION Dual Pump Packs, REF: OPO73
Quantity7644
DistributionUS: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
Initiated2022-12-07
Classified2023-01-20
Recall numberZ-1006-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Johnson & Johnson Surgical Vision, Inc. (Z-1006-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91290/

Not open yet

Alerts for Johnson & Johnson Surgical Vision, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.