Class II — LeMaitre Vascular, Inc.

Recalling firmLEMAITRE VASCULAR, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThree lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole, leading to the risk of the tip breaking off, and potentially becoming lost in the blood vessel.
ProductTufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire or introducer. The devices affected by this recall are 3F and 4F..
Quantity72 (OUS only)
DistributionNo US distribution. Foreign distribution to United Kingdom, Germany, France, Sweden, Austria, and Italy.
Initiated2022-09-30
Classified2022-11-11
Recall numberZ-0228-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — LeMaitre Vascular, Inc. (Z-0228-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91006/

Not open yet

Alerts for LEMAITRE VASCULAR, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.