Class II — Shanghai United Imaging Healthcare Co., Ltd.

Recalling firmShanghai United Imaging Healthcare Co.,Ltd.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThe wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.
ProductPositron Emission Tomography and Computed Tomography System, Model: uMI 550
Quantity32
DistributionDistribution to US states of TX, WI, CA, and MI
Initiated2022-05-30
Classified2022-10-13
Recall numberZ-0087-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Shanghai United Imaging Healthcare Co., Ltd. (Z-0087-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90888/

Not open yet

Alerts for Shanghai United Imaging Healthcare Co.,Ltd.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.