Class II — Siemens Healthcare Diagnostics, Inc.
| Recalling firm | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Dimension Vista® High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed. |
| Product | Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A |
| Quantity | 1592 units |
| Distribution | Nationwide Foreign: Canada |
| Initiated | 2022-08-24 |
| Classified | 2022-10-14 |
| Recall number | Z-0089-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-0089-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90872/
Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.