Class II — Angiodynamics, Inc.

Recalling firmANGIODYNAMICS, INC
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonOPS Fiber packaged within specific VenaCure EVLT Procedure Kits does not meet the required dimensional specification, may result in damage to the sheath, including separation or the inability to move the fiber within the sheath. A damaged sheath will necessitate removal from the patient, which may result in procedural delay and/or surgical intervention
ProductVenaCure EVLT KIT WITH SPOTLIGHT OPS SHEATH 55CM Ref./ Catalog No.: EVLT55OPS (each box contains 5 Procedure Kits). For use in the treatment of varicose veins.
Quantity40 units
DistributionUS Nationwide distribution in the states of FL, NC, NJ, NM, NY, PA, TX, VA.
Initiated2022-08-10
Classified2022-09-27
Recall numberZ-1819-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Angiodynamics, Inc. (Z-1819-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90811/

Not open yet

Alerts for ANGIODYNAMICS, INC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.