Class II — Merit Medical Systems, Inc.
| Recalling firm | Merit Medical Systems, Inc. |
|---|---|
| Registered facility | View this company's 15 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury. |
| Product | CorVocet Biopsy System/Catalog Code: CORA1610/B, CORA1610S/B, CORA1615/B, CORA1810/A, CORA1810/B, CORA1810S/B, CORA1815/A, CORA1815/B, CORA1815S/A, CORA1815S/B, CORA1820/A, CORA1820/B, CORA1820S/A, CORA1825/A, CORA1825/B, CORA1825S/A, CORA2010/A, CORA2010/B, CORA2010S/A, CORA2015/A, CORA2015/B, CORA2015S/A, CORA2020/A, CORA2020/B, CORCA1410/A, CORCA1410/B, CORCA1415/A, CORCA1415/B, CORCA1610/A, CORCA1610/B, CORCA1610S/B, CORCA1615/A, CORCA1615/B, CORCA1615S/B, CORCA1810B/A, CORCA1810B/B, CORCA1810SB/A, CORCA1810SB/B, CORCA1815B/A, CORCA1815B/B, CORCA1815SB/A, CORCA1815SB/B, CORCA1820B/A, CORCA… |
| Quantity | 1822 |
| Distribution | Worldwide distribution - US National distribution including in the states of FL, IL, OH, PA, CA, WI, VA, AZ, AR, MS, TN, MO, MA, MI, NY, NE, TX, WA, MN, LA, DE, IN, AL, NJ, OR, MD, GA, CT, OK, UT, KS, NC, CO, KY, WV, AK, NV, RI, SC, ND, NH, MT and the countries of Finland, Denmark, United Arab Emirates, Spain, Sweden, Norway, Jordan, Australia, Germany, Austria, Italy, United Kingdom, France, Cana… |
| Initiated | 2022-08-03 |
| Classified | 2022-10-27 |
| Recall number | Z-0137-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Merit Medical Systems, Inc. (Z-0137-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90804/
Alerts for Merit Medical Systems, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.