Class II — Qiagen Sciences LLC

Recalling firmQIAGEN Sciences LLC
Registered facilityQIAGEN (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDuring the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
ProductQIAcube Connect MDx, Model No. 9003070
Quantity93 US; 57 OUS
DistributionDomestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN, NV, NJ, NY, NC, OH, PA, SC, SD, TN, TX, VA, WA, WI. International distribution worldwide.
Initiated2021-11-21
Classified2022-09-15
Recall numberZ-1746-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Qiagen Sciences LLC (Z-1746-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90758/

Not open yet

Alerts for QIAGEN Sciences LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.