Class II — Qiagen Sciences LLC

Recalling firmQIAGEN Sciences LLC
Registered facilityQIAGEN (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIncreased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the patient being exposed to an inappropriate or suboptimal anti-cancer treatment and/or unnecessarily experiencing side effects
Producttherascreen FGFR RGQ RT-PCR kit (US IVD) REF 874721
Quantity22 kits
DistributionIN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands, Malaysia, Spain, Thailand
Initiated2022-07-20
Classified2022-08-29
Recall numberZ-1650-2022
InitiationVoluntary: Firm initiated