Class I — Synapse Biomedical Inc

Recalling firmSynapse Biomedical Inc.
Registered facilitySYNAPSE BIOMEDICAL INC. (the firm's only registered facility)
ClassificationClass I
StatusTerminated
ReasonDue to a defect identified on the EPG controller printed circuit board assembly, device performance may degrade over time. Depending on the failure mode, potential adverse events include pain, muscle fatigue and decreased breathing effectiveness, arrhythmia or cardiac arrest, or the device may stop unexpectedly.
ProductNeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021.
Quantity118 (112 US, 6 OUS)
Distributiondistribution to US states of AL, CO, IL, FL, GA, MD, MN, NC, NJ, OH, PA, and SC; Germany and Kuwait. ***Updated 3/28/23 - Product was not distributed to United Arab Emirates. Product was distributed to Kuwait.*** Updated 9.14.23 - 1 patient in Germany, 1 patient in the Czech Republic, 1 distributor in Kuwait
Initiated2022-07-11
Classified2022-08-22
Terminated2023-10-24
Recall numberZ-1561-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Synapse Biomedical Inc (Z-1561-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90626/

Not open yet

Alerts for Synapse Biomedical Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.