RegWardenFDA facility compliance records
Recall

Class II — Meridian Bioscience Inc

Recalling firmMERIDIAN BIOSCIENCE, INC.
Registered facilityMERIDIAN BIOSCIENCE, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonWhen a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.
ProductRevogene, Catalog no. 610210. IVD test instrument
Quantity459 (394 US, 65 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Chile, Italy, Qatar, and Switzerland.
Initiated2022-06-22
Classified2022-07-28
Recall numberZ-1479-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)