Class II — RAYSEARCH LABORATORIES AB

Recalling firmRaySearch Laboratories AB (publ)
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonAn issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
ProductRayStation 9B SP1. For radiation treatment planning.
Quantity1
DistributionWorldwide distribution - US Nationwide distribution in the state of TN and the countries of Belgium and Switzerland.
Initiated2022-06-01
Classified2022-07-19
Terminated2023-10-04
Recall numberZ-1403-2022
InitiationVoluntary: Firm initiated