Class III — Hamilton Medical AG
| Recalling firm | Hamilton Medical AG |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class III |
| Status | Completed |
| Reason | When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber. |
| Product | HAMILTON-H900 Humidifier, Models: 950001, 950004 |
| Quantity | 203 |
| Distribution | US: WI, FL, CA, NV, PA, TX, PA |
| Initiated | 2021-01-05 |
| Classified | 2022-07-14 |
| Recall number | Z-1385-2022 |
| Initiation | Voluntary: Firm initiated |